Z-1310-06: EDWARDS VIGILANCE MONITORS (FORMERLY BAXTER VIGILANCE MONITORS) with software release 5.3 or earlier only

Recall record Z-1310-06 for Edwards Lifesciences Llc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1310-06
Company in FDA record
Edwards Lifesciences Llc
FDA classification date
Aug 2, 2006
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Edwards Lifesciences Vigilance monitors with software release 5.3 or earlier may improperly cause the monitor to deliver power to the Continuous Cardiac Output (CCO) catheter without alerting the user to this situation. This can result in overheating and thermal damage to the CCO catheter and serious patient injury.

Source facts

Action
Strategy to contact affected physicians accounts: Edwards will communicate to all accounts with affected monitors using distribution and sales records to identify those accounts. Letter dated 07/07/2006. In addition Edwards will analyze all CCO catheter sales and identify all accounts not on the monitor distribution list and include those accounts on the communication listing. Communication will be send via FedEx to the Risk Manager at each account with sufficient copies for the Heads of Departments of Surgery, Anesthesia, ICU, Nursing, Risk Management and Biomedical Engineering. Edwards will follow up by phone with each account that does not initiate contact per the Recall Letter. Edwards will make three attempts to contact the account by phone and complete the recall process.
Initiated
2006-06-21
Root cause
Other
Terminated
2008-08-07
Distribution
Worldwide
Recall number
Z-1310-06
Recall status
Terminated
CDRH record ID
46857
FDA classified
2006-08-02
Recall event number
35777

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗