Z-1310-2019: MoniTorr 10100 10-100 CSF Drainage System w/Patient Line One Way Valve MoniTorr 10102 10-102 CSF Drainage

Recall record Z-1310-2019 for Integra LifeSciences Corp.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1310-2019
Company in FDA record
Integra LifeSciences Corp.
FDA classification date
May 20, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

The firm has identified a complaint trend regarding breakage of the LimiTorr Transducer Mount stopcock and MoniTorr Panel Mount stopcock.

Source facts

Action
On April 8, 2019, the firm distributed Urgent Medical Device Correction letters to its affected customers. The communication advised customers of the possible breakage of the stopcock. Customers were advised to take the following precautions: When connecting a fluid-filled transducer (FFT) to the Panel Mount/Transducer Mount stopcocks: " Do not apply excessive force while attaching the FFT to the stopcock. " Take care during the priming process to not torque or bend the syringe and FFT. " Ensure luer lock is "finger tight;" do not over-tighten connection at the stopcock junction. When managing an external ventricular or lumbar drain (EVD) with a FFT attached: " Take care when manipulating the stopcock to zero the FFT and to obtain pressure readings, do not apply excessive force to the stopcock junction. " Use caution during patient transportation and during routine ICU care, to avoid direct contact of the FFT with other devices or equipment. " If the FFT must be flushed or replaced, do not apply excessive torque or force to the FFT/ stopcock junction. " If the FFT is connected to the ICU bedside monitor, ensure the interface cable is positioned so that it does not place excess force on the FFT (e.Q. do not allow cable to touch the floor. Customers were asked to determine if they had any affected product on hand, review and understand the precautions, and complete and return a completed acknowledgement form to the firm. Distributors were asked to forward the customer letter to any customers who received the affected product. Should you have any questions regarding these instructions, please contact Customer Service Monday to Friday 8:00 AM - 8:00 PM EST USA: 1-800-654-2873: email: custsvcnj@integralife.com Outside USA: 781-565-1401 ***UPDATE*** On October 22, 2019, the firm sent out an updated notification informing customers that a new stopcock has been designed for the LimiTorr and MoniTorr, and is now available for replacement of th
Initiated
2019-04-08
Root cause
Device Design
Terminated
2021-01-27
Distribution
Worldwide Distribution - US Nationwide International distribution to Argentina, American Samoa, Australia, Canada, and Guam.
Recall number
Z-1310-2019
Recall status
Terminated
CDRH record ID
172225
FDA classified
2019-05-20
Recall event number
82659

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗