Z-1310-2023: Iradimed Corporation MRidium MRI 1000 Series Infusion Sets, MRI Infusuui Set - Syringe Adapter Set, REF 1

Recall record Z-1310-2023 for Iradimed Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1310-2023
Company in FDA record
Iradimed Corporation
FDA classification date
Mar 30, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Certain Iradimed 1057 Syringe Adapter Sets, used with the MRidium MRI infusion pump, have a problem relating to a defective injection molded part which can limit the effectiveness of the syringe venting function which then can result in occlusion.

Source facts

Action
Iradimed Corporation issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 02/24/2023 by certified mail. The notice explained the issue and the risk and provided updated instructions for use to mitigate the potential risk of inlet occlusion. For further information concerning this issue or have any questions, please contact Iradimed support via email at SyringeAdapterRecall@Iradimed.com or via telephone at 407-677-8022.
Initiated
2023-02-24
Root cause
Process control
Distribution
Worldwide distribution - US Nationwide and the countries of Austria, Canada, Chile, Denmark, Germany, Great Britain, Hong Kong, Indonesia, Italy, Korea, Kuwait, Malaysia, Netherlands, Nicaragua, Oman, Pakistan, Peru, Philippines, Poland, Portugal, Qatar, Romania, Serbia, Singapore, Singapore, South Africa, Spain, Switzerland, Taiwan, Thailand, United Arab Emirates.
Recall number
Z-1310-2023
Recall status
Open, Classified
CDRH record ID
199014
FDA classified
2023-03-30
Recall event number
91846

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗