Open original FDA record ↗Recall record
- FDA identifier
- Z-1311-2018
- Company in FDA record
- Roche Diagnostics Corporation
- FDA record creation date
- Apr 4, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Product exhibits a decreased recovery of patient samples on the cobas e 801 module compared to the MODULAR ANALYTICS E 170, cobas e 411, 601, and 602 analyzers.
Source facts
- Action
- The firm, Roche, sent an "URGENT MEDICAL DEVICE REMOVAL" letter dated December 22, 2017 to its customers. The letter described the product, problem and actions to be taken. The customers were instructed to do the following: " Immediately discontinue use of Elecsys BRAHMS PCT (catalog number 07301715190) reagent lot number 204084 on the cobas e 801 module. " Discard affected product according to local guidelines. " Complete the attached fax form and fax it to 1-800-722-7222. " File this Urgent Medical Device Removal (UMDR) for future reference. Please contact the Roche Support Network Customer Support Center, 24 hours a day, seven days a week at 1-800-428-2336 if you have questions about the information contained in this UMDR.
- Initiated
- 2017-12-22
- Root cause
- Software design
- Terminated
- 2019-04-05
- Distribution
- US Distribution in states of: CA, FL and IN.
- Recall number
- Z-1311-2018
- Recall status
- Terminated
- CDRH record ID
- 162386
- FDA record created
- 2018-04-04
- Recall event number
- 79462