Z-1311-2018: Elecsys BRAHMS PCT Immunoassay for the in vitro quantitative determination of PCT (procalcitonin) in hum

Recall record Z-1311-2018 for Roche Diagnostics Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1311-2018
Company in FDA record
Roche Diagnostics Corporation
FDA record creation date
Apr 4, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Product exhibits a decreased recovery of patient samples on the cobas e 801 module compared to the MODULAR ANALYTICS E 170, cobas e 411, 601, and 602 analyzers.

Source facts

Action
The firm, Roche, sent an "URGENT MEDICAL DEVICE REMOVAL" letter dated December 22, 2017 to its customers. The letter described the product, problem and actions to be taken. The customers were instructed to do the following: " Immediately discontinue use of Elecsys BRAHMS PCT (catalog number 07301715190) reagent lot number 204084 on the cobas e 801 module. " Discard affected product according to local guidelines. " Complete the attached fax form and fax it to 1-800-722-7222. " File this Urgent Medical Device Removal (UMDR) for future reference. Please contact the Roche Support Network Customer Support Center, 24 hours a day, seven days a week at 1-800-428-2336 if you have questions about the information contained in this UMDR.
Initiated
2017-12-22
Root cause
Software design
Terminated
2019-04-05
Distribution
US Distribution in states of: CA, FL and IN.
Recall number
Z-1311-2018
Recall status
Terminated
CDRH record ID
162386
FDA record created
2018-04-04
Recall event number
79462

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗