Skip to content

Public FDA record

Z-1311-2023: ARTIS pheno systems with a Siemens Healthineers table or a Trumpf/ MAQUET table-A Fluoroscopic X-Ray syst

Recall record Z-1311-2023 for Siemens Medical Solutions USA, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1311-2023
Company in FDA record
Siemens Medical Solutions USA, Inc
FDA classification date
Mar 30, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

In the event of any unintended table movement, the system may not detect the incorrect direction, could lead to the injury of a patient, staff member, operator, or equipment.

Source facts

Action
Siemens Healthcare issued A Customer Safety Advisory Notice distributed to customers on 2/2/23 via AX041/22/S. A software update to correct the error will be distributed to all affected customers via Update Instructions AX040/22/S. Affected systems will be updated to SW version VE10B. Letter states reason for recall, health risk and action to take: Please perform system movements with adequate care to avoid injuries to persons and damage to equipment due to collisions or crushing. Siemens will correct the software error via Update Instructions AX040/22/S. Our service organization will contact you shortly to arrange a date to perform this corrective action. Please feel free to contact our service organization for an earlier appointment at 1-800-888-7436. Please review, understand, and confirm acknowledgement of the information and instructions outlined in this Urgent Safety Advisory Notice. By signing this document, you are acknowledging that you have read and understand the content.
Initiated
2023-02-02
Root cause
Software design
Distribution
US Nationwide distribution.
Recall number
Z-1311-2023
Recall status
Open, Classified
CDRH record ID
198817
FDA classified
2023-03-30
Recall event number
91765

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗