Open original FDA record ↗Recall record
- FDA identifier
- Z-1311-2023
- Company in FDA record
- Siemens Medical Solutions USA, Inc
- FDA classification date
- Mar 30, 2023
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
In the event of any unintended table movement, the system may not detect the incorrect direction, could lead to the injury of a patient, staff member, operator, or equipment.
Source facts
- Action
- Siemens Healthcare issued A Customer Safety Advisory Notice distributed to customers on 2/2/23 via AX041/22/S. A software update to correct the error will be distributed to all affected customers via Update Instructions AX040/22/S. Affected systems will be updated to SW version VE10B. Letter states reason for recall, health risk and action to take: Please perform system movements with adequate care to avoid injuries to persons and damage to equipment due to collisions or crushing. Siemens will correct the software error via Update Instructions AX040/22/S. Our service organization will contact you shortly to arrange a date to perform this corrective action. Please feel free to contact our service organization for an earlier appointment at 1-800-888-7436. Please review, understand, and confirm acknowledgement of the information and instructions outlined in this Urgent Safety Advisory Notice. By signing this document, you are acknowledging that you have read and understand the content.
- Initiated
- 2023-02-02
- Root cause
- Software design
- Distribution
- US Nationwide distribution.
- Recall number
- Z-1311-2023
- Recall status
- Open, Classified
- CDRH record ID
- 198817
- FDA classified
- 2023-03-30
- Recall event number
- 91765