The stability specifications of ImmunoCAP Allergen c6, Amoxicilloyl were not met.
Source facts
Action
Customers are being asked to please stop using all current product, and scrap any unused products.
Please disregard any reported test result below the assay Limit of Quantification (LoQ 0.1 kUA/l) from the lots, as these results may not be valid.
Please note that assay results above the Limit of Quantification, will only be affected by giving a lower result than the real value.
Please fill in the Recall Letter Acknowledgement Form provided, including the table in the response form to obtain credit, and return the signed response by e-mail or fax.
Initiated
2017-12-18
Root cause
Device Design
Terminated
2019-03-27
Distribution
Nationwide Distribution
Recall number
Z-1313-2018
Recall status
Terminated
CDRH record ID
162339
FDA record created
2018-04-05
Recall event number
79444