Reason stated in the source
An electrical component within the system controller module of certain Bio-Console 560 devices may exhibit electrical failure causing the device to stop pumping. Issues observed include the pump stopping, the user interface going blank, and smoking and producing a burning odor during use.
Source facts
- Action
- In the United States (US), beginning 03-Mar-2021, a written Urgent Medical Device Correction letter was mailed via UPS 2-day delivery to consignees who have received affected product according to Medtronic records. Consignees will be asked to complete and return a confirmation certificate, confirming they have received the notification.
- Initiated
- 2021-03-03
- Root cause
- Component change control
- Distribution
- Worldwide distribution. US Nationwide including Puerto Rico, Brazil, Brunei Darussalam, Canada, China, Colombia, Croatia, Ecuador, Guatemala, India, Israel, Italy, Japan, Kazakhstan, Korea, Mexico, Morocco, New Zealand, Russian Federation, Spain, Taiwan, Thailand, United Kingdom, and Viet Nam
- Recall number
- Z-1313-2021
- Recall status
- Open, Classified
- CDRH record ID
- 186156
- FDA record created
- 2021-03-26
- Recall event number
- 87490