Skip to content

Public FDA record

Z-1313-2024: Steris Corporation Endo Cleaning Indicator VERIFY" RESITEST"

Recall record Z-1313-2024 for Mckesson Medical-Surgical Inc. Corporate Office. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1313-2024
Company in FDA record
Mckesson Medical-Surgical Inc. Corporate Office
FDA classification date
Mar 19, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Facility Critical Temperature Excursions on 08-JAN 2024 in Walk-In refrigerator due to mechanical failures, causing temperatures to fall below the validated range of 35.6 - 46.4 Degrees Fahrenheit (2 - 8 Degrees Celsius) for up to 60 minutes which could increase the probability of an incorrect test result or a delayed test result.

Source facts

Action
McKesson notified customers by phone January 9, 2024 with a follow-up call January 17, 2024 instructing customers of three scenarios based upon the products they received and the investigation in conjunction with relevant manufacturer. The firm also sent Urgent Product Recall Notifications via email on February 8, 2024 instructing customers to immediately examine inventory and quarantine all affected devices, complete and return the provided response form to MMSQRCProductReviews@Mckesson.com, and notify customers who you have further distributed recalled devices to of this recall and to segregate and quarantine any devices they have on hand. Replacement products will be provided and may have the same lot number(s) as the recalled items but are not affected by the recall and are acceptable for use. Customers with questions can contact McKesson directly by phone at (800) 688-8840. Representatives are available Monday to Friday, 8:00 am to 5:00 pm PST.
Initiated
2024-01-09
Root cause
Storage
Distribution
US Nationwide distribution in the states of AZ, PA and TX.
Recall number
Z-1313-2024
Recall status
Open, Classified
CDRH record ID
206161
FDA classified
2024-03-19
Recall event number
94024

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗