Skip to content

Public FDA record

Z-1315-06: Medplus ER Drug Screen Test Model #5800KAB packaged as 25 individually pouched devices in a labeled kit b

Recall record Z-1315-06 for Applied Biotech Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1315-06
Company in FDA record
Applied Biotech Inc
FDA classification date
Aug 1, 2006
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

The firm has discovered that a number of distributed devices of the Medplus ER Drug Screen Test may be incorrectly assembled which may give incorrect results. The product is an in vitro test for rapid detection of various drugs such as Amphetamines, Barbiturates, Phencyclidine, Cocaine, Methamphetamine, Tricyclic Antidepressants, Morphine, Tetrahydrocannabinol, & Benzodiazepines in human urine.

Source facts

Action
Recall notifications that describe the problem, dated May 31,2005, were sent via fax. Customers were instructed to block all stock of affected lots in their warehouse, contact their primary consignees and advise them to discontinue use immediately and dispose any unused materials. A faxback form was used to communicate to the recalling firm the quantity blocked and disposed of. Arrangements were made for providing replacement product.
Initiated
2005-05-31
Root cause
Other
Terminated
2012-07-25
Distribution
Nationwide (distributors in KS and OR).
Recall number
Z-1315-06
Recall status
Terminated
CDRH record ID
46564
FDA classified
2006-08-01
Recall event number
35656

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗