Open original FDA record ↗Recall record
- FDA identifier
- Z-1315-2008
- Company in FDA record
- Draeger Medical, Inc.
- FDA classification date
- Jul 16, 2008
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Detachment of Component: The chassis on some machines were manufactured out of specifications. As a result, the caster may break loose from the chassis.
Source facts
- Action
- Consignees were notified by an Urgent Medical Device Recall letter sent on 2/7/08. The letter informed users that the recalling firm will contact them to schedule appointments to inspect any potentially affected machines. The letter instructed users to continue using the machines, before the inspection, making sure to carefully follow the warnings in the instruction manual. If a caster on the machine appears unstable before the inspection is conducted, users were advised to contact the recalling firm. For additional information, contact 1-800-543-5047.
- Initiated
- 2008-02-05
- Root cause
- Nonconforming Material/Component
- Terminated
- 2008-09-09
- Distribution
- Nationwide Distribution.
- Recall number
- Z-1315-2008
- Recall status
- Terminated
- CDRH record ID
- 68298
- FDA classified
- 2008-07-16
- Recall event number
- 46806