Reference electrode used beyond the guaranteed in-use 52 week lifetime,may leak and potentially cause erroneous pH and/or sodium results. Other Ion parameters are not affected.
Source facts
Action
Roche Diagnostics Operations sent an Urgent Medical Device Correction letter dated February 27, 2014 via UPS Ground (receipt required) to all affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to complete the enclosed fax form and fax it to 1-888-345-0253. For questions contact Roche Customer Business Support Center at 1-800-428-5076.
Initiated
2014-02-28
Root cause
Device Design
Terminated
2018-01-10
Distribution
Worldwide Distribution - USA nationwide including PR and the country of Guam.
Recall number
Z-1315-2014
Recall status
Terminated
CDRH record ID
125702
FDA classified
2014-04-02
Recall event number
67549