The NBP measurement of Intellivue X3 Patient Monitor shows intermittently only mean values instead of the diastolic and systolic blood pressure values. This is caused by a falsely detected NBP cuff, which is leading to wrong internal NBP setting. Furthermore, occasionally the Monitor shows the Check Touch Input message and the monitor is inoperable with the touch interface.
Source facts
Action
Philips will contact customers with affected devices to arrange a correction of the issues. Firm plans a Software update of the affected devices and Replacement of the display assembly.
Initiated
2017-11-17
Root cause
Under Investigation by firm
Terminated
2020-09-16
Distribution
48 Foreign Accounts
Recall number
Z-1315-2018
Recall status
Terminated
CDRH record ID
160242
FDA classified
2018-04-06
Recall event number
78732