Skip to content

Public FDA record

Z-1315-2023: Color Cuff Single Port, Single Bladder 30 (Royal Blue) Quick Connect, REF: 5921-030-135; Color C

Recall record Z-1315-2023 for Stryker Sustainability Solutions. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1315-2023
Company in FDA record
Stryker Sustainability Solutions
FDA classification date
Apr 1, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Increased incidence of disposable tourniquet cuffs "not able to achieve desired pressure" and "not holding pressure" due to 1) Cuff fabric tearing during use, and 2) Pinched or unseated O-ring at the connection between the cuff, disposable fill line, and/or cuff hosing, which may lead to loss of vascular occlusion, which may lead to operative site blood loss, hemorrhage, or other complications.

Source facts

Action
On 3/1/23, recall notices started being distributed to customers. Customers were asked to do the following: 1) Follow the IFU and inspect for tears. Do not use the device if any damage is noted. Return the device and packaging to the recalling firm, if it is not in acceptable condition for surgery. 2) To avoid a pinched or unseated O-ring upon connection of the tourniquet cuff to the tourniquet controller, hold down the connector release on the tourniquet controller and take extra caution to ensure connection between the cuff and controller is not at an angle. 3) If a cuff does not reach or maintain the target pressure value during initial inflation, discontinue using the cuff. 4) Distribute the recall notice internally to all users. 5) If any of the affected devices have been forwarded to additional facilities, please contact these facilities and communicate this recall to them. Please list the additional facilities within the Business Reply Form 6) Following these steps, you may continue to use the product as-is. 7) Alternatively, you may return the impacted product. It is unclear when replacements will be available. 8) Complete and return the Recall Business Reply Form. It can be returned via email to SSSPFA@stryker.com Adverse reactions or quality problems experienced with the use of this product can be reported to the Complaint Hotline at 1-888-763-8803 or to http://www.stryker.com/productexperience Additional affected product was distributed, so additional recall notices were distributed to these associated customers beginning on 9/26/23.
Initiated
2023-03-01
Root cause
Process control
Distribution
US: WI, MD, NY, PA, WA, CO, GA, FL, OH, NC, TX, NJ, OR, MT, CA, RI, SC, NE, MI, LA, OK, SD, AL, VA, MS, WV, AZ, KY, IL, AR, ME, ND, MO, KS, HI, NH, AK, TN, IA, MN, IN, NV, MA, ID, DE, NM, WY, DC, VT, CT
Recall number
Z-1315-2023
Recall status
Open, Classified
CDRH record ID
199013
FDA classified
2023-04-01
Recall event number
91845

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗