Open original FDA record ↗Recall record
- FDA identifier
- Z-1316-06
- Company in FDA record
- Radiometer America Inc
- FDA classification date
- Aug 2, 2006
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The package insert, which accompanied the QC reagents used with ABL blood gas analyzers, referenced incorrect control ranges for Hct (hematocrit) testing. The range stated on the package insert was too low to generate accurate Hct test results (The control ranges were acceptable for the glucose and lactate testing).
Source facts
- Action
- The firm apprised their customers of this situation with a letter dated 5/24/2005. The letter requests that the customer remove the old (incorrect) package insert and replace it with a new (correct) insert which accompanied the customer letter. Should the customer have any questions, or need additional package inserts, the letter provides a toll-free contact telephone number.
- Initiated
- 2005-05-24
- Root cause
- Other
- Terminated
- 2012-02-24
- Distribution
- Nationwide, including: PA, MI, NY, OH, NJ, TX, AL, MO, AR, TN, WA, KY, and NC, and London, ON, Canada
- Recall number
- Z-1316-06
- Recall status
- Terminated
- CDRH record ID
- 46846
- FDA classified
- 2006-08-02
- Recall event number
- 35773