Z-1316-2013: Loop, Cutting, Angled 22FR (26055G/6). Resectoscope accessories (ie. electrode, cutting loops and cold k

Recall record Z-1316-2013 for Karl Storz Endoscopy America Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1316-2013
Company in FDA record
Karl Storz Endoscopy America Inc
FDA classification date
May 11, 2013
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Possible damage to the sterile packaging (blister tray) of some Karl Storz sterile products, which may result in airborne contamination.

Source facts

Action
A Safety Alert dated May 29, 2009 was sent to affected customers. The letter informed them of possible damage to the sterile packaging (blister tray) of some Karl Storz sterile products which may result in airborne contamination. Karl Storz requested consignees to inspect their inventory of the subject products thoroughly and look for small holes in the blister trays. Any identified units should be returned to Karl Storz, and a replacement package (6 units per package) will be provided to at no charge. The Karl Storz Customer Support number 1-800-421-0837 ext. 7577 was provided to obtain a Return Authorization Number (RMA) and arrange replacement. Questions and comments were forwarded to Technical Support at 1-800-421-0837 ext. 5350.
Initiated
2009-06-01
Root cause
Packaging
Terminated
2013-05-13
Distribution
Distributed Nationwide and in Canada.
Recall number
Z-1316-2013
Recall status
Terminated
CDRH record ID
90682
FDA classified
2013-05-11
Recall event number
55358

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗