Z-1316-2019: Abbott Alinity ci-series System Control Module software version 2.5.1; LN 3R70-01 with the following hard

Recall record Z-1316-2019 for Abbott Gmbh & Co. KG. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1316-2019
Company in FDA record
Abbott Gmbh & Co. KG
FDA record creation date
May 13, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Potential performance issues in the Alinity-ci software version 2.5.1

Source facts

Action
On March 7, 2019 Abbott issued the attached Product Correction letters to all installed Alinity ci series instrument customers: 1. Softwarerelated actions: Your Abbott representative will schedule a mandatory upgrade of your Alinity ciseries to software version 2.6.0 to resolve each of these issues. Refer to Appendix A for necessary actions required until software version 2.6.0 can be installed. Software version 2.6.0 also provides enhancements and additional changes to increase the usability of your system. If you require additional information, please contact Abbott Customer Service. 2. The Alinity ciseries Operations Manual distributed with software version 2.6.0 requires supplemental instructions for the following message: Message code 3696 The sample probe is not washed after aspirating a whole blood or red blood cell sample. If message code 3696 occurs, perform the following: a. Perform Asneeded Maintenance procedure 5906 Clean Sample and Reagent Probes (cseries) to clean the sample probe unless one of the following actions occurred before the initiation of assay processing: " The sample probe was replaced. " Daily maintenance procedure 5501 Daily Maintenance (cseries) was performed. b. Additional messages that are associated with this message can be used to determine the appropriate corrective action. Perform View additional messages that are associated with a message in the Alinity ciseries Operations Manual. 3. Bulk Solution Level Sensor related actions (refer to Appendix B for additional information): a. If LN 04S6801 Level Sensor, Bulk Solution is installed in the Alinity i Trigger or PreTrigger bulk solution reservoirs, Abbott recommends these sensors be replaced with a Level Sensor, Bulk Solution, LN 04S6802 before further test processing is performed. b. Abbott recommends all LN 04S6801 Level Sensor, Bulk Solution installed in the following bulk solution reservoirs be replaced with a Level Sensor, Bulk Solution, LN 04
Initiated
2019-03-07
Root cause
Under Investigation by firm
Terminated
2025-02-19
Distribution
AR, AZ, CA, FL, GA, IL, MA, ME, MN, MO, NY, OK, SC, TN, TX, UT, VA, WI, and Puerto Rico ALGERIA, ARGENTINA, AUSTRALIA, AUSTRIA, BAHRAIN, BANGLADESH, BELGIUM, BOSNIA AND HERZEGOVINA, BRAZIL, BULGARIA, CANADA, CHILE, CHINA, COLOMBIA, CROATIA, CZECH REPUBLIC, DENMARK, ESTONIA, FINLAND, FRANCE, GERMANY, GREECE, GUATEMALA, HONG KONG, INDONESIA, ISRAEL, ITALY, JAPAN, JORDAN, KAZAKHSTAN, KENYA, KUWAIT, LATVIA, LEBANON, MADAGASCAR, MALAYSIA, MEXICO, NETHERLANDS, NEW ZEALAND, NORWAY, PAKISTAN, PHILIPPINES, POLAND, PORTUGAL, ROMANIA, SAUDI ARABIA, SERBIA, SINGAPORE, SLOVAKIA, SOUTH AFRICA, SOUTH KOREA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, THAILAND, TRINIDAD & TOBAGO, TURKEY, UAE, UK, and VIETNAM.
Recall number
Z-1316-2019
Recall status
Terminated
CDRH record ID
171668
FDA record created
2019-05-13
Recall event number
82475

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗