Meter Temporarily Freezes-- A software compatibility issue may cause the blood glucose meter to cease operations and freeze temporarily.
Source facts
Action
The firm sent an Urgent Recall letter to Health Care Professionals on 12/18/2007. The letter advised user to uninstall software drivers and provided options on how to uninstall. For further assistance, please contact LifeScan Customer Service at 1-866-679-9250.
Initiated
2007-12-18
Root cause
Software design
Terminated
2008-07-18
Distribution
Nationwide Distribution including the states of AL, CA, IN, IO, MA, MD, MI, MS, NE, NY, NC, OR, PA, TN, TX, UT, VT, VA, WV, and WI.
Recall number
Z-1317-2008
Recall status
Terminated
CDRH record ID
67375
FDA classified
2008-04-08
Recall event number
46398