Open original FDA record ↗Recall record
- FDA identifier
- Z-1317-2013
- Company in FDA record
- ConMed Corporation
- FDA classification date
- May 13, 2013
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
ConMed was notified by the contract manufacturer that their wire/connector supplier reported damage to the connector tooling which could pinch or compress the lead wire filaments and potentially render the device incapable of delivering the appropriate energy or shock to the patient and/or could cause potential injury to user or patient.
Source facts
- Action
- ConMed sent a URGENT: Field Safety Notice letter dated May 3, 2013, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Customers were instructed to please review your inventory for any of the devices listed on Attachment I. Business Reply Forms were sent to the domestic and foreign consignees on May 3, 2013. The domestic notices were mailed by USP Priority Mail pouches. The recall is being conducted to the end user level. We ask that you contact all of those organizations within your facility and any other facilities that you may have supplied or given these affected products to. It is imperative that all end users of these devices receive this notice and respond immediately. If you have questions, please call 1 315-624-3237 or fax to +1 315-624-3225.
- Initiated
- 2013-05-03
- Root cause
- Component design/selection
- Terminated
- 2015-08-25
- Distribution
- Worldwide distribution: USA (nationwide) and the country of: Saudi Arabia.
- Recall number
- Z-1317-2013
- Recall status
- Terminated
- CDRH record ID
- 117902
- FDA classified
- 2013-05-13
- Recall event number
- 65112