- FDA identifier
- Z-1317-2015
- Company in FDA record
- Becton Dickinson & Company
- FDA classification date
- Mar 25, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Becton Dickinson received an elevated number of complaints for blood or infusate/medication leakage due to tearing of the silicone septum.
Source facts
- Action
- Becton Dickinson sent an Urgent Voluntary Product Recall Letter/Recall Response Card/Packing Instructions, dated 2/19/15, to their affected customers via UPS next day mail. The letters identified the affected device and explained the reason for the recall. Customers were to review their inventory for the specific catalog and lot numbers provided and quarantine the affected product. The Recall Response Card form should be completed and faxed back to BD at 1-866-514-2113 or e-mail bd7043@stericycle.com. All affected product should be returned with the completed Recall Response Card following the steps on the enclosed packing instruction. Upon receipt of the returned product, BD will send customers replacement product. If customers have questions or require assistance, they should contact 1-866-800-2917.
- Initiated
- 2015-02-19
- Root cause
- Process control
- Terminated
- 2017-05-25
- Distribution
- Distributed to the states of CA, CT, FL, IL, MA, ND, NY, OH and PA.
- Recall number
- Z-1317-2015
- Recall status
- Terminated
- CDRH record ID
- 134096
- FDA classified
- 2015-03-25
- Recall event number
- 70593