One lot of D-RAD Self-Taping Locking Screws used with the Distal Radius Fracture Kit were machined out of specification. Screws measured over tolerance within the head thread form by 0.002-0.003 inch.
Source facts
Action
The firm initiated their recall by letter and email on 03/07/2018. The firm requested return of the product.
Initiated
2018-03-07
Root cause
Under Investigation by firm
Terminated
2018-10-31
Distribution
US, Puerto Rico, Malaysia
Recall number
Z-1317-2018
Recall status
Terminated
CDRH record ID
162539
FDA classified
2018-03-20
Recall event number
79515
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.
Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.