Skip to content

Public FDA record

Z-1317-2025: Brand Name: ZOLL Product Name: Powerheart G5 AED Model/Catalog Number: (1) G5A-80A, (2) G5A-80A-TSO. Soft

Recall record Z-1317-2025 for ZOLL Medical Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1317-2025
Company in FDA record
ZOLL Medical Corporation
FDA classification date
Mar 12, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The potential for devices to fail their self-test as a result of prolonged exposure to extreme environments (temperature and/or humidity).

Source facts

Action
On February 13, 2025, URGENT: MEDICAL DEVICE CORRECTION letters were sent to customers. REQUIRED ACTIONS Customers who have AEDs should routinely: (1) Monitor device for self-test results. (2) For customers who experience a self-test failure, take action to resolve any alerts as soon as possible. Customers who have at-risk devices should immediately: (1) Communicate this notification to G5 AED users. (2) Ensure users have a program in place that periodically monitors the status of these devices, in accordance with G5 User s Guide. (3) EC 501 is likely due to exposure to humidity outside the device s published specification. Ensure devices are stored within the device s published specifications: " Temperature: 0C to 50C (32F to 122F) " Humidity: 10% to 95% (non-condensing) (4) Respond to ZOLL via the customer notification form acknowledging that you have received this notice. We have notified the appropriate regulatory agencies of this correction and expect it to be classified as a recall. We apologize for any inconvenience this may cause you and thank you in advance for assistance in implementing this correction. Ensuring these devices are rescue ready is our highest priority. Our 24/7 technical support numbers 1 (800) 348-9011 or 1 (978) 421-9460 and are available to assist users with any aspect of this notice. Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail or by fax with the fax number being 1-800-FDA-0178.
Initiated
2025-02-13
Root cause
Unknown/Undetermined by firm
Distribution
Domestic: Nationwide Distribution International: AE, AI, AM, AR, AT, AU, BD, BE, BG, BH, BM, BN, BS, CA, CH, CL, CN, CO, CR, CZ, DE, DK, DO, EG, ES, FI, FR, GB, GH, GU, HK, HR, HU, ID, IE, IL, IN, IS, IT, JE, JM, JP, KR, KW, KZ, LB, LT, MA, MC, MD, ME, MQ, MS, MT, MX, MY, NC, NG, NL, NO, NZ, OM, PA, PE, PG, PH, PK, PL, PR, PT, QA, RE, RO, RS, RU, SA, SE, SG, SI, SK, SM, TC, TH, TN, TR, TW, UA, US, VG, ZA.
Recall number
Z-1317-2025
Recall status
Open, Classified
CDRH record ID
212403
FDA classified
2025-03-12
Recall event number
96198

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗