Z-1318-2010: Access Immunoassay Systems Accu Tnl Reagent Kit when used in conjunction with UniCel Dxl Access Immunoass

Recall record Z-1318-2010 for Beckman Coulter Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1318-2010
Company in FDA record
Beckman Coulter Inc.
FDA classification date
Apr 12, 2010
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

The recall was initiated because Beckman Coulter has confirmed customer reports that different results have been obtained using the same patient samples on Access/Access 2 and UniCel Dxl platforms. Values obtained with UniCel Dxl systems have been demonstrated to have a positive bias compared to values obtained with Access or Access 2 systems.

Source facts

Action
An "URGENT: PRODUCT CORRECTIVE ACTION (PCA)" letter dated February 5, 2010, was sent to the customers. The letter describes the product, problem and action to be taken by the customers. The customers are recommended to immediately discontinue using AccuTnI Reagents and Calibrators on the UniCel DxI system until further notice. If this is not possible, the customers are directed to notify their stakeholders of the bias until further notice. The customers should complete, fax (786-639-4000) or mail the enclosed RESPONSE FORM. If consignees need assistance or have any questions regarding the notification, they were instructed to contact Technical Support at 1-800-854-3633 in the United States and Canada.
Initiated
2010-02-06
Root cause
Other
Terminated
2012-05-09
Distribution
Worldwide distribution: USA and Canada
Recall number
Z-1318-2010
Recall status
Terminated
CDRH record ID
89075
FDA classified
2010-04-12
Recall event number
54729

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗