Open original FDA record ↗Recall record
- FDA identifier
- Z-1318-2010
- Company in FDA record
- Beckman Coulter Inc.
- FDA classification date
- Apr 12, 2010
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The recall was initiated because Beckman Coulter has confirmed customer reports that different results have been obtained using the same patient samples on Access/Access 2 and UniCel Dxl platforms. Values obtained with UniCel Dxl systems have been demonstrated to have a positive bias compared to values obtained with Access or Access 2 systems.
Source facts
- Action
- An "URGENT: PRODUCT CORRECTIVE ACTION (PCA)" letter dated February 5, 2010, was sent to the customers. The letter describes the product, problem and action to be taken by the customers. The customers are recommended to immediately discontinue using AccuTnI Reagents and Calibrators on the UniCel DxI system until further notice. If this is not possible, the customers are directed to notify their stakeholders of the bias until further notice. The customers should complete, fax (786-639-4000) or mail the enclosed RESPONSE FORM. If consignees need assistance or have any questions regarding the notification, they were instructed to contact Technical Support at 1-800-854-3633 in the United States and Canada.
- Initiated
- 2010-02-06
- Root cause
- Other
- Terminated
- 2012-05-09
- Distribution
- Worldwide distribution: USA and Canada
- Recall number
- Z-1318-2010
- Recall status
- Terminated
- CDRH record ID
- 89075
- FDA classified
- 2010-04-12
- Recall event number
- 54729