Skip to content

Public FDA record

Z-1318-2012: Anesthesia Machines (digital display assembly); A gas machine for anesthesia is a device used to admi

Recall record Z-1318-2012 for Penlon, Ltd.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1318-2012
Company in FDA record
Penlon, Ltd.
FDA classification date
Mar 26, 2012
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Penlon, Ltd has identified the potential for the premature failure of a component in some Electronic Flowmeter sensor display unit PCB assemblies.

Source facts

Action
Penlon Limited sent a Technical Bulletin dated November 22, 2011, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. As a precaution a field upgrade part was released that required all existing PCB assemblies to be removed and the updated item installed. The old PCB must be returned to Penlon with the serial number of the affected machine. The required replacement parts were included with the technical bulletin. For further information customers should contact the Technical Support Team at techsupport@penlon.com.
Initiated
2011-11-29
Root cause
Component design/selection
Terminated
2012-07-03
Distribution
Worldwide Distribution - USA (nationwide) and the countries of Peru, Uganda, Kenya, Malaysia, Poland, Saudi Arabia, United Arab Emirates, Philippines, Equador, and Thailand
Recall number
Z-1318-2012
Recall status
Terminated
CDRH record ID
106127
FDA classified
2012-03-26
Recall event number
60647

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗