Reason stated in the source
Penlon, Ltd has identified the potential for the premature failure of a component in some Electronic Flowmeter sensor display unit PCB assemblies.
Source facts
- Action
- Penlon Limited sent a Technical Bulletin dated November 22, 2011, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. As a precaution a field upgrade part was released that required all existing PCB assemblies to be removed and the updated item installed. The old PCB must be returned to Penlon with the serial number of the affected machine. The required replacement parts were included with the technical bulletin. For further information customers should contact the Technical Support Team at techsupport@penlon.com.
- Initiated
- 2011-11-29
- Root cause
- Component design/selection
- Terminated
- 2012-07-03
- Distribution
- Worldwide Distribution - USA (nationwide) and the countries of Peru, Uganda, Kenya, Malaysia, Poland, Saudi Arabia, United Arab Emirates, Philippines, Equador, and Thailand
- Recall number
- Z-1318-2012
- Recall status
- Terminated
- CDRH record ID
- 106127
- FDA classified
- 2012-03-26
- Recall event number
- 60647