Open original FDA record ↗Recall record
- FDA identifier
- Z-1318-2013
- Company in FDA record
- GE Healthcare, LLC
- FDA classification date
- May 14, 2013
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
While performing a fluoroscopic examination on the GE Innova systems, there is a potential of nonrecoverable loss of imaging modes with no advance warning, due to a relay failure in the power distribution box (PDB). Although no injuries have been reported due to this issue to date, there is risk of complications during the procedure as a result of imaging functionality loss.
Source facts
- Action
- GE Healthcare did not send out customer letters. GE Healthcare completed the correction by replacing the faulty relay in all affected systems. As an effectiveness check, they have tracked all service requests for the field action to completion. These service requests recorded the repair of the affected units. Exceptions to completion of the service request were recorded as well. Records of field action completion will be retained per their processes and procedures. Should you have any questions or require additional information, please call (262) 548-2757.
- Initiated
- 2011-12-20
- Root cause
- Nonconforming Material/Component
- Terminated
- 2014-03-13
- Distribution
- Worldwide Distribution -- Nationwide including the states of AL, AZ, AR, CA, CO, CT, DC, FL, GA, GUAM, HI, ID, IL , IN, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, MT, NE, NV, NH, NH, NM, NY, NC, ND, OH, OK, OR, PA, PR, SC, TN, TX, UT, VT, VA, WA, WY and WI., and the countries of Philippines, Peru, Paraguay, Panama, Mexico, Malaysia, Korea, Japan, Guatemala, Ecuador, Dominican Republic , Columbia, China, Chile, Canada, Brazil.,Argentina and Venezuela
- Recall number
- Z-1318-2013
- Recall status
- Terminated
- CDRH record ID
- 117691
- FDA classified
- 2013-05-14
- Recall event number
- 65029