Z-1318-2023: COVIDIEN Shiley Lo-Pro Oral/Nasal Tracheal Tube Cuffed REF 86051 7.0 mm I.D. The Shiley Lo-Pro Oral/Nasa

Recall record Z-1318-2023 for Covidien. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1318-2023
Company in FDA record
Covidien
FDA classification date
Apr 1, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to potential leaks associate with the Oral/Nasal Tracheal Tube Cuffed inflation system.

Source facts

Action
On 02/10/2023, Medtronic issued an "Urgent Medical Device Recall" letter via UPS 2 day delivery to customers informing them of potential cuff leaks from the device inflation system in Shiley Lo-Pro Oral/Nasal Tracheal Tube Cuffed. Customer are instructed to: 1. Quarantine all unused product from the affected lots of Shiley" Lo-Pro Oral/Nasal Tracheal Tube Cuffed. See attachment A for guidance on identifying potentially affected devices. 2. Return all unused product from the affected lots in their inventory to Medtronic as described on the Customer Confirmation Form. 3. Complete the enclosed Customer Acknowledgment Form even if they do not have unused inventory. 4. Pass on the notice to all those who need to be aware within their organization or to any organization where the potentially affected product from the specified lots has been transferred or distributed. Additional Information for Patient Management: As stated in the Instructions for Use that accompanies each device, each tube s cuff should be tested by inflation prior to use. Use of a device with this potential manufacturing assembly error may result in a delay to treatment while an alternate device is obtained. There are no additional patient management recommendations that should be employed for patients where potentially affected devices are currently in use or were used. These patients should be monitored in accordance with your medical facility s critical care protocols. Clinical staff should appropriately assess and manage patients for any adverse clinical outcomes. For questions or further assistance, contact a Medtronic Representative or Customer Service at 800-962-9888.
Initiated
2023-02-10
Root cause
Process control
Distribution
U.S.: AK, AZ, CA, FL, GA, IA, IL, IN, KS MD, MI, MS, NC, NE, NJ, NY, OH, OK, PA, SC, TX, UT, VA, WA, and WV O.U.S.: Canada
Recall number
Z-1318-2023
Recall status
Open, Classified
CDRH record ID
198983
FDA classified
2023-04-01
Recall event number
91841

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗