Skip to content

Public FDA record

Z-1318-2024: Assembly, Array Guide TMINI, REF: 109101, part of the TMINI Knee Instrument Set, used with the TMINI Mini

Recall record Z-1318-2024 for Think Surgical, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1318-2024
Company in FDA record
Think Surgical, Inc.
FDA classification date
Mar 20, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There is a galling effect (cold-weld) caused by friction between the array fixation pins and the TMINI array guide holes due to insufficient clearance between the outer diameter of the fixation pins and the inner diameter of the TMINI array guide holes. As a result, the array guide has the possibility of binding to the array fixation pins during drilling.

Source facts

Action
On 2/12/24, recall notices were mailed and emailed to customers who were asked to do the following: 1) Examine your inventory, locate any unused devices and quarantine them immediately. Your clinical representative(s) will assist with the recall process and provide guidance to your surgeon(s) for positioning the array fixation pins using the array bracket. 2) Do not use any affected product and return any used or unused product. 3) Complete and return the acknowledgment form to productreturns@thinksurgical.com Customers with questions can contact the firm via phone at 510-249-2300, Option 1.
Initiated
2024-02-12
Root cause
Device Design
Distribution
US Nationwide distribution in the states of AL, OH, NV, TX, UT, CA, MO.
Recall number
Z-1318-2024
Recall status
Open, Classified
CDRH record ID
206295
FDA classified
2024-03-20
Recall event number
94086

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗