Z-1318-2026: Spectral CT 7500 on Rails; Product Code (REF): 728334;

Recall record Z-1318-2026 for Philips North America Llc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1318-2026
Company in FDA record
Philips North America Llc
FDA classification date
Feb 6, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

As part of the CT systems sub assembly, thirty-two (32) fasteners are used to attach the rotor to the bearing within the gantry. One or more of these fasteners may not be torqued to specification. If multiple fasteners are not torqued to specification, the rotor or rotor parts may become unsecured or displaced. While there has been no reported or observed event of expelled parts during gantry rotation, unsecured rotor parts may potentially be expelled.

Source facts

Action
On or about January 7, 2026 URGENT Medical Device Correction letters were sent to customers. Actions to be taken: 1. As a precaution, monitor the gantry for any unusual sounds or vibrations during operation. If unusual sounds are heard or vibrations are observed, immediately call your service contact. In the absence of unusual sounds or vibrations, you may continue to use your system(s) according to the intended use. 2. Circulate this Urgent Medical Device Correction Letter to all users of this device so that they are aware of the issue. Please retain this letter with your system(s) until Philips has inspected your system and has provided any correction, if needed. Ensure the letter is in a place likely to be seen and read. Actions planned by Philips: Philips will contact you to schedule a time for a Philips Field Service Engineer (FSE) to visit your site and to inspect the fasteners on your system (reference FCO72800845). If the inspection indicates that your system is affected, Philips will perform the correction to resolve the issue (reference FCO72800846)
Initiated
2026-01-07
Root cause
Process change control
Distribution
Worldwide - US Nationwide distribution in the states of AL, AR, CA, CO, CT, FL, GA, HI, IL, IN, KY, LA, MA, MD, MN, NC, NY, OH, OR, PA, TX, VT, WA, WV and the countries of Australia, Austria, Bahrain, Belgium, Brazil, Bulgaria, Chile, China, Costa Rica, Croatia, Czech Republic, Denmark, Ecuador, Egypt, El Salvador, France, Germany, Hong Kong, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kuwait, Lithuania, Macao, Mexico, Netherlands, New Zealand, Norway, Panama, Philippines, Poland, Portugal, R¿union, Russian Fed., Saudi Arabia, Serbia, Singapore, South Africa, South Korea, Spain, Switzerland, Taiwan, Thailand, Turkmenistan, United Kingdom, Uzbekistan, Vietnam.
Recall number
Z-1318-2026
Recall status
Open, Classified
CDRH record ID
218016
FDA classified
2026-02-06
Recall event number
98248

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗