Skip to content

Public FDA record

Z-1319-2021: EPIX Therapeutics DiamondTemp Bidrectional Small Curve Ablation Catheter, 7.5 French Size(2.48 mm), Conn

Recall record Z-1319-2021 for EPIX THERAPEUTICS, INC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1319-2021
Company in FDA record
EPIX THERAPEUTICS, INC
FDA record creation date
Mar 29, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

It has been identified that the catheter size on the packaging labeling does not correspond to the widest part of the catheter. In addition the IFU contains the incorrect diameter of the catheter. This could result in blood stagnation, clots or air emboli.

Source facts

Action
On February 22, 2021, Medtronic issued a "Urgent Medical Device Correction" notification via regionally approved methods to consignees who have received affected product. In addition to informing consignees about the recall the firm asked consignees to take the following actions: 1. Review the corrected language applicable to IFU and label as provided in the Appendix below. Until the IFU is updated with the corrected information, the information provided in the Appendix should be used in combination with the existing labeling provided with the DiamondTemp" Catheters. 2. Please share this information with healthcare professionals in your facility that use DiamondTemp" Catheters. Also share this information with any other organization where these devices may have been transferred. 3. Please maintain a copy of this notice in your records. We are committed to patient safety and welcome any questions you may have regarding this communication. Please contact your Medtronic representative with questions.
Initiated
2021-02-22
Root cause
Under Investigation by firm
Distribution
Product was not distributed in the US. OUS: Germany, Spain, France, Belgium, United Kingdom, Slovenia, Italy, Canada
Recall number
Z-1319-2021
Recall status
Open, Classified
CDRH record ID
185985
FDA record created
2021-03-29
Recall event number
87412

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗