Open original FDA record ↗Recall record
- FDA identifier
- Z-1320-06
- Company in FDA record
- Radiometer America Inc
- FDA classification date
- Aug 3, 2006
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The dimensions and shape of the analyzer adaptor do not fit the analyzer tubes or probe, causing air to be aspirated with the blood sample and causing a deviation in the pO2 sample results.
Source facts
- Action
- The firm issued a letter to their customers dated 6/9/2005. The letter reports the defect with the ABL5 Capillary Adapters and informs the customers that Radiometer reps will be replacing all of the customers'' ABL5 adaptors ASAP. The letter is accompanied by a single bag of new adaptors for the customers'' ''immediate needs'' and also provides a faxed response form which the customer can use to request additional adapters as required. In addition, should the customer choose to use the old adapters in their possession, the letter stresses and reiterates the ''Precautions and Warnings'' applicable for use of the ABL5 should air bubbles be observed in the measuring chamber by the user/operator.
- Initiated
- 2005-06-09
- Root cause
- Other
- Terminated
- 2012-02-24
- Distribution
- Nationwide, including KS, OH, MD, FL, TX, CA, KY, MA, VA, WA, IL, AR and ME, and Canada.
- Recall number
- Z-1320-06
- Recall status
- Terminated
- CDRH record ID
- 46847
- FDA classified
- 2006-08-03
- Recall event number
- 35774