Open original FDA record ↗Recall record
- FDA identifier
- Z-1320-2026
- Company in FDA record
- Medline Industries, LP
- FDA classification date
- Feb 6, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class I
Reason stated in the source
These lots of reprocessed catheters may contain small particles of residual material. If affected devices are used, there is a risk of inflammatory response, systemic infection, or granulomatous reaction or thrombus reaction which may lead to cerebral or pulmonary embolism, or deep vein thrombosis. Expansion of recall Z-2610/2614-2025
Source facts
- Action
- On December 22, 2025, the firm issued Urgent Medical Device Recall letters as a follow-up to an earlier action in 2025 for the same issue, but newly identified item numbers and lots of previously identified item numbers. Notably, the products associated with the new/updated action are to be destroyed by the customer on-site for credit, and not returned to the manufacturer. ***Updated 5/18/26*** The firm issued an expanded notification on 5/15/26 which expanded the action to include on additional lot of product. These devices were not expired at the time of the original notification but have since expired.
- Initiated
- 2025-12-22
- Root cause
- Under Investigation by firm
- Distribution
- Distribution US nationwide.
- Recall number
- Z-1320-2026
- Recall status
- Open, Classified
- CDRH record ID
- 218102
- FDA classified
- 2026-02-06
- Recall event number
- 98277