Certain lots of M-30D diluent used with BC-3200 and BC-3600 Hematology Analyzers and M-53D diluent used with BC-5390 Hematology Analyzer may cause the system to produce an elevated platelet background count.
Source facts
Action
Letters dated January 24, 2018, were sent to consignees with the following instructions: The issue manifests intermittently, therefore, you may continue to use affected lots if the background check is successfully completed. Discard affected lots which cause the system to display the error message background abnormal. Mindray will replace the potentially affected lots at no cost. Please contact Technical Support to order replacement product. For further questions, please call (201) 995-8407.
Initiated
2018-01-24
Root cause
Material/Component Contamination
Terminated
2019-04-19
Distribution
US Distribution including Puerto Rico and to the states of : CA, MI, NC, TX, and WA. .
Recall number
Z-1321-2018
Recall status
Terminated
CDRH record ID
162631
FDA record created
2018-04-09
Recall event number
79556