Open original FDA record ↗Recall record
- FDA identifier
- Z-1321-2021
- Company in FDA record
- EPIX THERAPEUTICS, INC
- FDA record creation date
- Mar 29, 2021
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
It has been identified that the catheter size on the packaging labeling does not correspond to the widest part of the catheter. In addition the IFU contains the incorrect diameter of the catheter. This could result in blood stagnation, clots or air emboli.
Source facts
- Action
- On February 22, 2021, Medtronic issued a "Urgent Medical Device Correction" notification via regionally approved methods to consignees who have received affected product. In addition to informing consignees about the recall the firm asked consignees to take the following actions: 1. Review the corrected language applicable to IFU and label as provided in the Appendix below. Until the IFU is updated with the corrected information, the information provided in the Appendix should be used in combination with the existing labeling provided with the DiamondTemp" Catheters. 2. Please share this information with healthcare professionals in your facility that use DiamondTemp" Catheters. Also share this information with any other organization where these devices may have been transferred. 3. Please maintain a copy of this notice in your records. We are committed to patient safety and welcome any questions you may have regarding this communication. Please contact your Medtronic representative with questions.
- Initiated
- 2021-02-22
- Root cause
- Under Investigation by firm
- Distribution
- Product was not distributed in the US. OUS: Germany, Spain, France, Belgium, United Kingdom, Slovenia, Italy, Canada
- Recall number
- Z-1321-2021
- Recall status
- Open, Classified
- CDRH record ID
- 185987
- FDA record created
- 2021-03-29
- Recall event number
- 87412