Z-1321-2023: Tempus Pro Patient Monitor is a portable vital signs monitor intended to be used by clinicians and medica

Recall record Z-1321-2023 for Remote Diagnostic Technologies Ltd.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1321-2023
Company in FDA record
Remote Diagnostic Technologies Ltd.
FDA classification date
Apr 1, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

During internal testing, Regulatory Compliance issues were identified regarding the Monitor and Power Supplies for fluid ingress and basic safety issues.

Source facts

Action
On 02/17/2023, the firm sent an "URGENT Medical Device Correction" Letter via priority mail to customers informing them that during internal testing, Regulatory Compliance issues regarding Fluid Ingress and Basic Safety issues were identified with the Tempus Pro Monitor, AC Mains Power Supply, and the Vehicle Adaptor. Customers are instructed to: " As there is currently no indication that any of the issues have resulted in a degradation in product performance, Philips advises that the Tempus Pro devices should remain in service. " Continue to follow the User/Operator Manual, including: o Take care to ensure that water or liquids are not spilt over the device or into its ventilation holes in the side corners. o Check connector covers, particularly the capnometer door cover, to ensure they close and latch acceptably. " Post this Urgent Medical Device Correction letter on or near your Tempus Pro device. " Complete and return the Urgent Medical Device Correction response form included, no later than 30 days from receipt. Philips will be providing a new and compliant AC Mains Power Supply to all affected customers. A new and compliant Vehicle Adaptor will also be provided to affected customers with devices that include that option. Philips anticipates providing this solution to customers in the second half of 2023. For questions or further information, contact the local Philips Representative, or Philip at (800)263-3342.
Initiated
2023-02-23
Root cause
Under Investigation by firm
Distribution
Worldwide Distribution: U.S.(nationwide): AL, AZ, CA, CO, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, TN, TX, UT, VA, WA, WI, WV, WY and O.U.S.(foreign): United Arab Emirates, Afghanistan, Austria, Australia, Belgium, Brunei Darussalam, Bolivia, Canada, Switzerland, Czechia, Germany, Denmark, Spain, France, United Kingdom, Hong Kong, Ireland, Israel, Iceland, Italy, Cayman Islands, Luxembourg, Malaysia, Netherlands, Norway, Portugal, Romania, Saudi Arabia, Sweden, Singapore, Slovenia, Thailand, Mayotte, and South Africa.
Recall number
Z-1321-2023
Recall status
Open, Classified
CDRH record ID
198918
FDA classified
2023-04-01
Recall event number
91805

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗