Reason stated in the source

Due to a manufacturing issue, the screw seat location on the abutment product may be too high, causing mating screw to have a shorter engagement length into the associated implant.

Source facts

Action
On 02/09/2024, the firm sent an "URGENT: MEDICAL DEVICE RECALL" Letter via US certified mail to customers to inform them that the abutment was manufactured with a dimensional condition that may lead to reduced engagement between the prosthetic screw and the implant. Customer are instructed to inspect their inventory and discard or return the affected products to PREAT CORPORATION. For questions or further assistance contact the Quality Assurance Manager at 800-232-7732 or email at nfetterman@younginnovations.com On 2/16/2024, the firm sent via certified USPS mail an "URGENT: MEDICAL DEVICE RECALL UPDATE" which informs customer to notify all personnel within their company, and of their customers where the affected products may have been further distributed of the recall.
Initiated
2024-02-09
Root cause
Process change control
Distribution
U.S. Nationwide distribution in the states of AZ, OH and PA.
Recall number
Z-1321-2024
Recall status
Open, Classified
CDRH record ID
206173
FDA classified
2024-03-20
Recall event number
94032

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗