Open original FDA record ↗Recall record
- FDA identifier
- Z-1322-06
- Company in FDA record
- Radiometer America Inc
- FDA classification date
- Aug 3, 2006
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Capilliary end caps, an accessory for ABL blood gas analyzers, leak blood from capillary tubes during transport.
Source facts
- Action
- The firm notified their customers of this recall action with a letter dated 3/22/2006. The letter was accompanied by a single bag of new end caps. The letter informs the customers of the ''leakage'' problem and asks that the customer examine their inventories and discard any of the referenced end caps found on-hand at any of their various locations. The letter also includes a response form to be faxed back to Radiometer indicating the amount of replacement capillary end caps required by the customer in addition to the one bag received. The recalling firm also notified the customers that the suspect end caps will all be replaced ASAP.
- Initiated
- 2006-03-22
- Root cause
- Other
- Terminated
- 2012-02-24
- Distribution
- Nationwide, Canada and Istanbul.
- Recall number
- Z-1322-06
- Recall status
- Terminated
- CDRH record ID
- 46927
- FDA classified
- 2006-08-03
- Recall event number
- 35775