Open original FDA record ↗Recall record
- FDA identifier
- Z-1322-2022
- Company in FDA record
- Covidien, LP
- FDA record creation date
- Jul 1, 2022
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The affected staplers have the potential for the staple guide to not be securely attached to the instrument. The issue could cause the component to disengage and if disengaged, could allow the device to transect tissue without forming staples.
Source facts
- Action
- Beginning 13-MAY-2022 communications to all affected consignees began. In the United States notifications are set via UPS 2-day delivery. Customers should return all unused affected product in inventory to Medtronic. Consignees are asked to forward this notification to those who need to be aware within the organization and to any location which the devices have been transferred. Consignees are asked to complete and return a confirmation certificate (or equivalent record) confirming they have received the notification.
- Initiated
- 2022-05-13
- Root cause
- Process change control
- Distribution
- Worldwide distribution - US Nationwide.
- Recall number
- Z-1322-2022
- Recall status
- Open, Classified
- CDRH record ID
- 194232
- FDA record created
- 2022-07-01
- Recall event number
- 90362