The recalling firm was notified by the contract manufacturer that they had reversed the L2, L3, L4, or L5 components on the P/N09010 RoHS boards of the Coblator II Controller (Model RF8000E).
Source facts
Action
ArthroCare sent an Urgent Medical Device Recall letter dated February 2015 to customers. A response form will be included in the letter and a follow up notifcation letter will be sent to non-responding consignees. The letter identified the affected product, problem and actions to be taken. For questions contact your local Returns Department Representative at 800-343-5717, option 3.
Initiated
2015-02-20
Root cause
Packaging process control
Terminated
2016-02-26
Distribution
US Nationwide Distribution in the states of TN, IA, AZ, NE, SD, NY, and OK.
Recall number
Z-1323-2015
Recall status
Terminated
CDRH record ID
134100
FDA classified
2015-03-26
Recall event number
70594