Skip to content

Public FDA record

Z-1323-2022: EIT CIF cage, H 6mm, 8degree, L- an intervertebral body fusion devices indicated for use with autograft w

Recall record Z-1323-2022 for DePuy Spine, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1323-2022
Company in FDA record
DePuy Spine, Inc.
FDA record creation date
Jul 1, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Label on the Outer Carton does not match the labels on the Outer Blister package, Inner Blister package and Patient Labels. The only label that may not be correct is on the Outer Carton

Source facts

Action
DePuy Synthes Spine issued Urgent Medical Device Recall Letter dated 5/18/21 to hospitals, sales consultants and sales managers. Letter states reason for recall, health risk and action to take: 1. Examine your inventory immediately to determine if you have the subject products and quarantine the subject products. DO NOT USE THE SUBJECT PRODUCTS. 2. Contact your DePuy Synthes Spine Sales Consultant to coordinate the return/credits of the subject products. 3. Review, complete, sign, and return the attached Business Response Form (See page 3 of this communication) to OneMD-Field-Actions@its.jnj.com within three (3) business days of receipt of this notification. Please include FA 2113300: 2022 EIT Cage in the e-mail subject line. IMPORTANT: Please complete the attached Business Response Form even if you do not have any units of the subject product on hand. 4. Forward this notice to anyone in your facility that needs to be informed (i.e., those who manage, transport, store, stock, or use the subject products). 5. If any of the subject product has been forwarded to another facility, contact that facility, and provide them with a copy of this notice. 6. Post a copy of this notice in a visible area for awareness and keep a copy for your record. Contact your DePuy Synthes Spine Sales Consultant. For Medical Information Requests, please visit our website: https://www.jnjmedicaldevices.com/mir
Initiated
2022-05-20
Root cause
Labeling Change Control
Distribution
Distribution US Nationwide
Recall number
Z-1323-2022
Recall status
Open, Classified
CDRH record ID
194284
FDA record created
2022-07-01
Recall event number
90390

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗