Z-1324-2013: Segmented Cylinder Applicator Set, Model # GM11004150, when used for Pulsed Dose Rate (PDR) Product Usag

Recall record Z-1324-2013 for Varian Medical Systems, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1324-2013
Company in FDA record
Varian Medical Systems, Inc.
FDA classification date
May 15, 2013
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Segmented cylinder applicator set may slip during treatment, causing the delivery of radiation to areas outside the target.

Source facts

Action
Varian Medical Systems, Inc. notified customers of the recall with 'Urgent Medical Device Correction Urgent Field Safety Notice," dated 4/11/13, and delivered by mail to customers. The notice identified the product, problem, and instructions for the customers. Varian Medical Systems, Inc. informed customers of the possible malfunction of the GM11004150 Segmented Cylinder Applicator Set for pulsed dose rate brachytherapy treatment, with codes H640383, H640277, H640384, H640788, H640D126, H640385, H640372, H640303, H640387. The notice instructed customers to cease use of the Segmented Cylinder Applicator Set and return the devices along with a completed proof of notification form (included with the recall notice) to the recalling firm.
Initiated
0013-04-11
Root cause
Device Design
Terminated
2014-01-13
Distribution
Worldwide distribution: USA (nationwide) and in the countries of: Belgium, France, Martinique, Slovenia, Sweden, and the United Kingdom.
Recall number
Z-1324-2013
Recall status
Terminated
CDRH record ID
117584
FDA classified
2013-05-15
Recall event number
64986

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗