Open original FDA record ↗Recall record
- FDA identifier
- Z-1324-2013
- Company in FDA record
- Varian Medical Systems, Inc.
- FDA classification date
- May 15, 2013
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Segmented cylinder applicator set may slip during treatment, causing the delivery of radiation to areas outside the target.
Source facts
- Action
- Varian Medical Systems, Inc. notified customers of the recall with 'Urgent Medical Device Correction Urgent Field Safety Notice," dated 4/11/13, and delivered by mail to customers. The notice identified the product, problem, and instructions for the customers. Varian Medical Systems, Inc. informed customers of the possible malfunction of the GM11004150 Segmented Cylinder Applicator Set for pulsed dose rate brachytherapy treatment, with codes H640383, H640277, H640384, H640788, H640D126, H640385, H640372, H640303, H640387. The notice instructed customers to cease use of the Segmented Cylinder Applicator Set and return the devices along with a completed proof of notification form (included with the recall notice) to the recalling firm.
- Initiated
- 0013-04-11
- Root cause
- Device Design
- Terminated
- 2014-01-13
- Distribution
- Worldwide distribution: USA (nationwide) and in the countries of: Belgium, France, Martinique, Slovenia, Sweden, and the United Kingdom.
- Recall number
- Z-1324-2013
- Recall status
- Terminated
- CDRH record ID
- 117584
- FDA classified
- 2013-05-15
- Recall event number
- 64986