Open original FDA record ↗Recall record
- FDA identifier
- Z-1324-2015
- Company in FDA record
- TITAN SPINE, LLC
- FDA classification date
- Mar 26, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The TCS Straight Hexlobular #6 Drivers (5210-1004) tip sheared off during use in procedures that involved patients with dense bones. As a result of this malfunction, Titan Spine, LLC is removing all of the TCS Straight Hex lobular #6 Drivers (5210-1004) Lot# G150101 from the field.
Source facts
- Action
- Titan Spine sent an "Urgent Medical Device Recall" letter via Fed Ex dated February 20, 2015, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. A replacement TCS Straight Hex lobular# 6 Drivers (5210-1004) will be shipped out to each Titan Spine Area Vice President and Sales Rep. within 4-6 weeks. For further questions please call (262) 242-7801.
- Initiated
- 2015-02-20
- Root cause
- Nonconforming Material/Component
- Terminated
- 2015-04-17
- Distribution
- US Distribution to the states of : AZ, CA, CT, FL, GA, ID, IL, KY, MD, MO, NV, NY, OH, OK, OR, TX, UT and WI.
- Recall number
- Z-1324-2015
- Recall status
- Terminated
- CDRH record ID
- 134071
- FDA classified
- 2015-03-26
- Recall event number
- 70576