Open original FDA record ↗Recall record
- FDA identifier
- Z-1325-2009
- Company in FDA record
- MDS Canada Inc. DBA MDS Nordion
- FDA classification date
- Jun 5, 2009
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Instructions for the use of the pinch clamp that is an integral part of the TheraSphere Administration Set were not included with the distribution of the device.
Source facts
- Action
- A User Bulletin (TS-2009-1) dated February 23, 2009 was issued to all TheraSphere user sites. The bulletin described the issue, affected units, and recommended actions for users. The bulletin instructed users on the proper use of the pinch clamp. MDS Nordion also updated the package insert with these instructions. For questions about the recall, customers in North America contact MDS Nordion Customer Service at 1-866-363-3330 and customers outside North America call 1-877-474-5999 and request Customer Service. Customers may also e-mail MDS Nordion at TheraSphere@mdsinc.com.
- Initiated
- 2009-02-24
- Root cause
- Error in labeling
- Terminated
- 2009-06-08
- Distribution
- Worldwide Distribution -- US (states of AZ, CA, CO, FL, IL, IN, KS, MD, MI, NC, RI, OH, OR, PA, WI) and country of Canada.
- Recall number
- Z-1325-2009
- Recall status
- Terminated
- CDRH record ID
- 81031
- FDA classified
- 2009-06-05
- Recall event number
- 51701