Skip to content

Public FDA record

Z-1325-2014: 5/10 mm Weck Vista Optical Bladeless Laparoscopic Access Port, Catalog No. 405910R; 5/10/12 mm Weck Vista

Recall record Z-1325-2014 for Teleflex Medical. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1325-2014
Company in FDA record
Teleflex Medical
FDA classification date
Apr 2, 2014
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Complaints of leakage of insufflation gas through the device.

Source facts

Action
Letters were issued on March 11, 2014 to customers informing them of the recall. The letter asked the consignee to immediately discontinue use and quarantine any products on hand. The letter included a Recall Acknowledgement Form which is to be returned. A letter was also issued to distributors on March 11, 2014, requesting them to conduct a sub-recall.
Initiated
2014-03-11
Root cause
Nonconforming Material/Component
Terminated
2016-01-28
Distribution
Worldwide Distribution-USA (nationwide) including Puerto Rico and the states of AL, AZ, AR, CA, CO, CT, FL, GA, IL, IN, KS, KY, MD, MA, MI, MN, MS, MO, NH, NJ, NY, NC, ND, OH, OR, PA, SC, TN, UT, VA, WA, WV, WI and the countries of Australia, Brazil, Canada, China, Israel, Korea, Malaysia, Singapore, and Vietnam.
Recall number
Z-1325-2014
Recall status
Terminated
CDRH record ID
126432
FDA classified
2014-04-02
Recall event number
67819

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗