Added Warning Statement: Fresenius 2008 Series Hemodialysis Machines False Blood Leak Alarm When Dialyzing Patients Treated with Hydroxocobalamin (or any form of Vitamin B-12)
Source facts
Action
Fresenius Medical Care issued an Urgent Medical Device Alert dated March 10, 2016, to all affected consignees. Consignees were instructed to place notification and Warning Addendum with their 2008 Series Hemodialysis Machine Operators Manual.
A reply form was included as part of the customer communication with instructions to fill out and return via fax or scanned/emailed copy.
Further information or support concerning this issue, customers were instructed to contact Fresenius Technical Services at 800-227-2572.
Initiated
2016-03-10
Root cause
Device Design
Terminated
2017-03-09
Distribution
Nationwide Distribution
Recall number
Z-1325-2016
Recall status
Terminated
CDRH record ID
144284
FDA classified
2016-04-05
Recall event number
73487