Skip to content

Public FDA record

Z-1326-2020: TLC-2000 Therapeutic Medical Laser System - Product Usage: Indicated for the temporary relief of knee pai

Recall record Z-1326-2020 for Theralase Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1326-2020
Company in FDA record
Theralase Inc.
FDA classification date
Feb 13, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class III

Reason stated in the source

Power supply (Model Number: TLC-2003) ME90A2403F01 (24 VDC, 3.75 A, 90 W, 1% peak to peak ripple and noise, which meets UL/EN/IEC-60601-1-2, 4th edition specification) shipped versus Power supply (Model Number: TLC-2003) MENB1100A2403F01 (24 VDC, 4.2 A, 101 W, 1% peak to peak ripple and noise, which meets UL/EN/IEC-60601-1-2, 3rd edition specification)

Source facts

Action
Email was sent out to healthcare practitioners regarding the recall on September 26, 2019. Healthcare practitioners are instructed to courier their TLC-2000 to Theralase. A Field Inspection Test will be completed and once passed, a Special Inspections for Medical Electrical Equipment and Systems label will be applied to the TLC-2000.The Field Inspection Test will be completed according to either CSA 60601-1 or UL 60601-1 dependent upon the customers Canadian or United States location, respectively.
Initiated
2019-09-26
Root cause
Process control
Terminated
2020-12-03
Distribution
Worldwide distribution - US Nationwide and countries Australia, Bahrain, Canada, Hong Kong Province of China, Jordan, Mexico, Pakistan, Philippines, Thailand, Romania.
Recall number
Z-1326-2020
Recall status
Terminated
CDRH record ID
179531
FDA classified
2020-02-13
Recall event number
84845

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗