Z-1326-2021: Philips SureSigns VM4, VM6 and VM8 - Product Usage: patient monitors-for monitoring, recording and alarmi

Recall record Z-1326-2021 for Philips North America, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1326-2021
Company in FDA record
Philips North America, LLC
FDA record creation date
Mar 30, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Fail to Comply with Chinese Standard YY1079-2008: Clauses 4.2.6 and 4.2.7.3. Range/accuracy of heart rate meter for pediatric mode-In pediatric mode, when the input signal rate is over 300 bpm, the indicated rate of the affected products will be lower than this upper limit

Source facts

Action
Philips Market Organizations are responsible for distributing the letters outside of the U.S. Letter issued on 3/23/21 , stating: Reason for correction, health risk and Action to be Taken by Customer/User states: This issue only happens on pediatric type patient with heartrate is greater than device claimed maximum heart rate 300 bpm. The device can continue to be used, however when using the monitor in the Pediatric category, patients at risk for tachycardia should be kept under close observation in accordance with existing product warnings described in the IFU. You may also consider to keep the default HR alarm limits or set tight HR alarm limits around the patients baseline HR for these patients Philip will send an engineer onsite to check the SW revision of your monitors. PHILIPS Subsequently a SW upgrade will be performed if your device is with the affected software revision. A Philips Healthcare representative will contact you to arrange for the correction.
Initiated
2021-03-23
Root cause
Software design
Terminated
2023-09-28
Distribution
International distribution in the country of China.
Recall number
Z-1326-2021
Recall status
Terminated
CDRH record ID
185993
FDA record created
2021-03-30
Recall event number
87418

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗