Generator Failure resulting increase of catheter temperature that could not be controlled by power reduction.
Source facts
Action
Customers were instructed to suspend use of the generators with software version 1.033 until they are upgraded with version 1.034B in a letter dated 7/15/2004. Firm issued a second letter on 9/17/04 for a software error discovered in-house during the upgrade. It instructs a workaround solution to prevent unwanted activity of the catheter.
Initiated
2004-07-16
Root cause
Other
Terminated
2012-05-07
Distribution
Nationwide
Recall number
Z-1328-04
Recall status
Terminated
CDRH record ID
34180
FDA classified
2004-08-07
Recall event number
29591
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.
Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.