Skip to content

Public FDA record

Z-1328-2013: Iris Diagnostics Division iQ Control/Focus Set, Part Number 800-3104, Lot Number 153-12. Product Usage:

Recall record Z-1328-2013 for Iris Diagnostics. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1328-2013
Company in FDA record
Iris Diagnostics
FDA classification date
May 15, 2013
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The firm initiated this recall because complaints from customers reported that the recovery compared to assigned assay values is significantly below specification. This results in the instrument not passing controls and the customer not being able to run patient samples.

Source facts

Action
IRIS sent a Medical Device Recall letter dated September 18, 2012 to all affected customers. The letter identified the affected product, problem, and actions to be taken. Customers were instructed to dispose of all unopened or partially used inventory, complete the attached worksheet certifying the disposal of the affected product and email or fax to Iris Diagnostics.
Initiated
2012-09-18
Root cause
Nonconforming Material/Component
Terminated
2014-01-27
Distribution
Worldwide Distribution - USA Nationwide and the countries of: Canada, Malaysia, Spain, Switzerland, Taiwan, Peru, France, Russia, Japan, United Kingdom, Germany, and Paraguay.
Recall number
Z-1328-2013
Recall status
Terminated
CDRH record ID
117370
FDA classified
2013-05-15
Recall event number
64910

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗