Open original FDA record ↗Recall record
- FDA identifier
- Z-1328-2013
- Company in FDA record
- Iris Diagnostics
- FDA classification date
- May 15, 2013
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The firm initiated this recall because complaints from customers reported that the recovery compared to assigned assay values is significantly below specification. This results in the instrument not passing controls and the customer not being able to run patient samples.
Source facts
- Action
- IRIS sent a Medical Device Recall letter dated September 18, 2012 to all affected customers. The letter identified the affected product, problem, and actions to be taken. Customers were instructed to dispose of all unopened or partially used inventory, complete the attached worksheet certifying the disposal of the affected product and email or fax to Iris Diagnostics.
- Initiated
- 2012-09-18
- Root cause
- Nonconforming Material/Component
- Terminated
- 2014-01-27
- Distribution
- Worldwide Distribution - USA Nationwide and the countries of: Canada, Malaysia, Spain, Switzerland, Taiwan, Peru, France, Russia, Japan, United Kingdom, Germany, and Paraguay.
- Recall number
- Z-1328-2013
- Recall status
- Terminated
- CDRH record ID
- 117370
- FDA classified
- 2013-05-15
- Recall event number
- 64910