Skip to content

Public FDA record

Z-1328-2021: Alcon Custom Ophthalmic Surgical Procedure Packs

Recall record Z-1328-2021 for Alcon Research, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1328-2021
Company in FDA record
Alcon Research, LLC
FDA classification date
Apr 1, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Reports have been received concerning mold found on gowns included in custom ophthalmic surgery packages.

Source facts

Action
On February 2, 2021 Alcon issued an :Urgent: Voluntary Medical Device Removal letter via FedEx mail to all affected consignees. On February 12, 2021 and April 9, 2021, Alcon expanded their recall to additional lots and consignees. On June 14, 2021, Alcon expanded their recall and issued an URGENT: VOLUNTARY MEDICAL DEVICE FIELD CORRECTION (FIELD ACTION EXPANSION) letter to affected consignee via FedEx. In addition to informing consignees about the recall, the firm asked consignees to take the following actions: 1. Forward this notification to all departments or organizations using your Custom Pak. 2. See list (Table 1) of affected Alcon Custom Pak lots that were or will be distributed to your facility. You may receive Custom Pak lots assembled prior to this Field Correction notification, and as a result, you may be receiving affected inventory in future shipments. If you receive any affected Alcon Custom Pak lots in the future, they will be identified with the one of the following stickers: A. "ATTENTION DISCARD ENCLOSED GOWNS REPLACE WITH INDIVIDUALLY WRAPPED STERILE GOWSNS" or B. "URGENT FIELD CORRECTION Discard Enclosed Gowns Replace with individually wrapped sterile gowns" 3. Review your inventory of Alcon Custom Pak lots against the attached table. For Custom Pak in your inventory, affix the provided stickers directly to the outside of your identified affected Custom Pak units in a location most easily seen for your facility. NOTE: Alcon Custom Pak units and standalone gowns are sterile and should not be opened prior to surgery. 4. Upon opening of your identified Alcon Custom Pak for surgical use, remove and dispose of the enclosed surgical gowns. 5. Use separate sterile standalone surgical gowns, either your own inventory or those supplied by Alcon. 6. Please complete the attached Response Form indicating your understanding of the included instructions. 7. Return the attached Response Form via fax or email to Alcon. 8. Keep Table 1 for your rec
Initiated
2021-02-02
Root cause
Nonconforming Material/Component
Terminated
2023-06-30
Distribution
Distribution US nationwide, including Puerto Rico; Canada, Australia, New Zealand, Mexico, Malaysia, and Thailand
Recall number
Z-1328-2021
Recall status
Terminated
CDRH record ID
185790
FDA classified
2021-04-01
Recall event number
87307

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.