Open original FDA record ↗Recall record
- FDA identifier
- Z-1329-04
- Company in FDA record
- Ev3
- FDA classification date
- Aug 7, 2004
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Ev3 had received customer complaints indicating that the stent device could not be used with an .035' guidewire. An investigation found that 6Fr, 135 cm (.018') stent system were labeled as a 120 cm (.035') stent system and that 6Fr 120 cm (.035') were labeled as 6Fr, 135 cm (.018').
Source facts
- Action
- Send notification letter to direct accounts and an EV3 representative will contact the direct account within the next few days to arrrange for the return of all unused product.
- Initiated
- 2004-07-09
- Root cause
- Other
- Terminated
- 2006-12-28
- Distribution
- Eleven (11) devices were distributed to seven (7) U.S. customers. Ten (10) devices were distributed to seven (7) clinical study sites in the U.S.
- Recall number
- Z-1329-04
- Recall status
- Terminated
- CDRH record ID
- 34246
- FDA classified
- 2004-08-07
- Recall event number
- 29629