Skip to content

Public FDA record

Z-1329-2024: HDS Traction Boot II, REF AR-6529-23, The Traction Boot is designed to position, support and/or distract

Recall record Z-1329-2024 for Baxter Healthcare Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1329-2024
Company in FDA record
Baxter Healthcare Corporation
FDA classification date
Mar 20, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Baxter Healthcare Corporation is issuing an Urgent Medical Device Recall for the Hip Distraction System, Traction Boot II, due to the boots being assembled incorrectly. The locking ridge on the front face of the boot key socket is facing upwards, which prevents the socket mechanism on the traction boot from attaching properly to the spar of the traction system. This will cause the boot to detach from the traction system.

Source facts

Action
Baxter issued an Urgent Medical Device Recall notice to its sole consignee on 03/11/2024 via email. The notice explained the issue and potential hazard and requested the following: 1. Immediately locate, isolate, and cease all use of the affected serial numbers of the product. The product code and serial number can be found on labelling on the bottom of the boot. 2. Arrange for the return of your impacted product by calling the Baxter Order Management Team at 800-433-5774, option 2 then 3, between the hours of 8:00 am and 5:00 pm Eastern Time, Monday through Friday, or by emailing Baxter at hrc_amatech_orders@baxter.com. 3. Acknowledge the receipt of this notification by completing a reply form on the customer portal.
Initiated
2024-03-11
Root cause
Under Investigation by firm
Distribution
US Nationwide distribution in the state of Florida.
Recall number
Z-1329-2024
Recall status
Open, Classified
CDRH record ID
206559
FDA classified
2024-03-20
Recall event number
94202

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗